Endress+Hauser Analog pH sensor Orbipore CPS91
Glass electrode for heavily polluted media in chemical processes, paper or paint production
Description
Specifications
| Measuring principle | Potentiometric | |
| Application |
Process applications |
media with high contamination: emulsions, suspensions, precipitation reactions
High levels of electrode poisons, such as H2S
|
| Min. conductivity |
>500µS/cm
|
|
| Characteristic | rapid response – new, stabilised gel insensitive to contaminations | |
| Measurement range | pH 0-14 | |
| Measuring principle | Open aperture – double gel reference | |
| Design |
all shaft lengths with temperature sensor – advanced gel technology
|
|
| Material | Glass | |
| Dimension | Diameter | 12 mm (0.46 inch) |
| Shaft lengths | 120, 225, 360 and 425 mm (4.68, 8.77, 14.04 and 16,57 inch) |
|
| Process temperature | Max. | 110°C (230°F) |
| Process pressure |
Max. | 10 bar(145 psi) |
| Temperature sensor |
Optional with integrated Pt100 or Pt1000. | |
| Ex certification | Simple operational equipment, no certification necessary. | |
| Connection | TOP68 plug-in head | |
| Ingress protection | IP68 | |