Endress+Hauser Digital combined pH/ORP sensor Memosens CPS16E

Memosens 2.0 glass electrode for standard applications in process and water & wastewater industries

Category:
Description

Specifications

Measuring principle Potentiometric
Application Long-term monitoring and limit control in processes with stable process conditions
Chemical processes
Pulp and paper industry
Waste incinerators and gas scrubbers
Mining
Wastewater
Characteristic Combination of pH and ORP electrode for standard applications in process technology and environmental engineering
Poison-resistant reference with ion trap
Measurement range ORP –1 500 to 1 500 mV
Application B pH: 0 to 14
Application A pH: 0 to 12
Measuring principle Gel compact electrode with PTFE ring-junction and with ion trap
Pt-disc as additional ORP element
rH measurement and control of the reference impedance
Design
All shaft lengths with temperature sensor
Advanced gel technology
Material Sensor shaft Glass to suit process
pH membrane glass Type A, B
Metal lead Ag/AgCl
Open aperture PFTE junction, sterilizable
ORP measuring element Platinum
O-ring FKM
Process coupling PPS fiber-glass reinforced
Nameplate Ceramic metal oxide
Dimension Diameter 12 mm (0.46 inch)
Shaft lengths 120, 225 and 360 mm
(4.72, 8.86, 14.17 and 16.73 inch)
Process temperature Application A  –15 to 80 °C (5 to 176 °F)
Application B 0 to 135 °C (32 to 275 °F)
Process pressure
0.8 to 17 bar (11.6 to 246,5 psi) absolute
Temperature sensor
NTC 30k
Ex certification With ATEX, IECEx, CSA C/US, NEPSI, Japan Ex and INMETRO approvals for use in hazardous areas Zone 0, Zone 1 and Zone 2
Connection Inductive, digital connection head with Memosens 2.0 technology
Ingress protection IP68

Download