Endress+Hauser Hydrostatic Level measurement Deltapilot FMB52
Pressure sensor with Contite measuring cell for hydrostatic level measurement
Description
Pressure sensor with Contite measuring cell for hydrostatic level measurement
Specifications
Characteristic | Pressure transmitter for hydrostatic level measurement with flush mounted metallic Contite measuring cell: Hermetically sealed, condensate-resistant, climatic-proofed and with lowest temperature influences Cable version |
Supply voltage | 4…20 mA HART 10,5…45 VDC (Non Ex): Ex ia: 10,5…30 VDC PROFIBUS PA: 9…32 VDC (Non Ex) FOUNDATION Fieldbus: 9…32 VDC (Non Ex) |
Reference Accuracy | Standard 0.2% Optional 0.1% |
Long term stability | 0.05 % of URL/year |
Process temperature | PE cable: -10°C…70°C / 14°F…158°F FEP cable: -10°C…80°C / 14°F…176°F |
Ambient temperature | -40°C…+85°C (-40°F…+185°F) |
Measuring cell | 100 mbar…10 bar (1.5 psi…150 psi) |
Smallest calibratable span | 100 mbar…10 bar (1.5 psi…150 psi) |
Vacuum resistance | 0 mbar abs. |
Max. Turn down | 100:1 |
Max. overpressure limit | 40 bar (600 psi) |
Process connection | Thread: G1 1/2, MNPT1 1/2 Flange: DN40…DN100, ASME 2″…4″, JIS 10K |
Material process membrane | 316L, AlloyC, Gold-Rhodium PE, FEP |
Material gasket | Viton, EPDM, Kalrez, none |
Fill fluid | Inert oil, Synthetic oil |
Material housing | 316L, Die-cast aluminum |
Communication | 4…20 mA HART PROFIBUS PA FOUNDATION Fieldbus |
Certificates / Approvals | ATEX, FM, CSA, CSA C/US, IEC Ex, INMETRO, NEPSI, UK Ex |
Safety approvals | Overfill protection WHG SIL |
Design approvals | EN10204-3.1 NACE MR0175 |
Marine approvals | GL/ ABS/ LR/ BV/ DNV |
Drinking water approvals | KTW/ NSF/ ACS |
Specialities | Modularity to differential pressure and process pressure devices (replacable display, universal electronics) Diagnostic functions |