METTLER pH Sensor InPro4260/SG/225/PT1000
The InPro 4260 SG is an analog, in-line combined pH and ORP (redox) probe with integrated temperature sensor designed for use in highly demanding chemical applications.
Reliable, Long-Lasting pH Probe.
The InPro 4260 SG is an analog, in-line combined pH and ORP (redox) probe with integrated temperature sensor designed for use in highly demanding chemical applications.
Withstands Demanding Applications
The InPro 4260 SG uses a polymer reference system for precise pH measurement and long lifetime, even under difficult industrial conditions.
Accurate pH & ORP Measurement
The InPro 4260 SG provides a wide measurement range from 0-14 pH and has a solution ground for redox measurement and enhanced stability.
Designed for Reliability
The InPro 4260 SG has a domed glass membrane that impedes bubble formation and an integrated temperature sensor to ensure reliable measurement.
ISM (Digital) or Analog | Analog |
PH Sensing Technology | Highly Alkali Resistant Glass Membrane for Tough Conditions (HA) |
PH Range | 0 – 14 |
Reference System | Xerolyt Extra polymer electrolyte, open double junction, Argenthal cartridge with silver-ion trap |
Operating Temperature (Fahrenheit) | 32 °F – 266 °F |
Operating Temperature (Celsius) | 0 °C – 130 °C |
Cable Connector | VP |
Process Connection | Pg 13.5 Thread |
Sterilizable/Autoclavable | No |
Sensor Length | 225 mm |
Certificates & Approvals |
CE
METTLER TOLEDO Quality Certificate
FM: IS Cl. I, II, III, Div 1, GR ABCDEFG/T6
ATEX/IECEx: Ex ia IIC T6/T5/T4/T3 Ga/Gb
PED
|
Operating Pressure (Bar) |
0 to 15 barg @ 25°C
0 to 7 barg @ 130°C
|
Model Family |
InPro4260(i)
|
Temperature Sensor |
PT 1000
|
Measurement Parameter |
pH
ORP/Redox
|
Shaft Material |
Glass
|
Operating Pressure (PSI) |
0 to 101 psig @ 266°F
0 to 217 psig @ 77°F |
Solution Ground/Redox Metal | Platinum |
Refillable Electrolyte |
No
|
Hazardous Area Approvals |
Yes
|
Primary Media/Application |
Chemical Processes
|
Hygienic/Biocompatibility |
No
|