MKS 910 DualTrans™ MicroPirani™ – Absolute Piezo Vacuum Pressure Transducers

Absolute Measurement Range: 1.0 x 10-5 to 1,500 Torr

Category:
Description

Specifications

Absolute Measurement Range 1.0 x 10-5 to 1,500 Torr
Digital Communication RS485/RS232 (4800-230400 Baud)
Sensor
Type 1: MicroPirani (MEMS Thermal Conductivity)
Type 2: Piezo Absolute (MEMS Diaphragm)
Relays 3 relays SPDT (optional)
Installation Orientation Any
Set Point Range Absolute: 5.0 x 10-4 to 1,500 Torr
Calibration Gas Air, Argon, Helium, Nitrogen, H2, H2O vapor, CO2, xenon, Neon
Operating Temperature 0-40°C (32-104°F)
Maximum Bakeout Temperature 85°C (185°F), non-operating
Digital Controls Zero adjust, span adjust, analog output, pressure units, baud rate, address, factory default, set point functions: value, hysteresis, direction, enable, transducer status, switch, LED test
Digital Status Pressure reading, units, set point, operating time, transducer temperature, user tag, model, device type, serial number, firmware and hardware versions, part number, manufacturer
Analog Output 1-9.2 VDC, 1 VDC/decade, 100Ω maximum output impedance (absolute pressure)
Analog Output Resolution 16 bit
Relay Contact Rating 1 A @ 30 VAC/VDC resistive
Relay Response
100 msec maximum
Power Requirements
9-30 VDC, 150 ma, < 1.2 watts maximum
Accuracy
5 x 10-4 to 1 x 10-3 Torr ±10% of reading
1 x 10-3 to 11 Torr ±5% of reading
11 to 1000 Torr ±0.75% of reading
Repeatability
5 x 10-4 to 10-3 Torr ±8% of reading
10-3 Torr to 11 Torr ±2% of reading
11 to 1000 Torr ±0.2% of reading
Overpressure Limit
2250 Torr (absolute)
Internal Volume
2.8 cc (KF16)
Exposed Materials
Silicon, SiO2, Si3N4, gold, low outgassing epoxy resin, 304 stainless steel, Viton®
Electronic Casing and Flange
304 Stainless Steel
Weight
0.37 lb (170 g) with KF16 flange
Compliance
CE

Download